Online Odor Monitor

102 

The vigi e-nose is the first analyzer able to track VOC & Sulfurs at ppb/ppt levels. It is more sensitive than human nose. It is composed by one autoGC-ED (MEDOR® Electrochemical wet cell Detector) and PID for VOCs. Zero air generator included in the same box for autonomous operation.

Standard version: H2S / SO2 / MM / EM / DMS / DMDS / DES / Total VOC
Upon request: TBM / NPM / MES / 2BM / IBM / NBM / THT / IPM

 

Principle:

  • Automatic sampling using a loop linked to sampling pump
  • Automatic loop injection on metallic capillary column
  • Isothermal gas chromatograph
  • Detection of all compounds eluting from the column performed by MEDOR® Detector: Electrochemical wet cell Detector which is SSD Sulfur Specific Detector.
  • Signal provided by reaction between the wet cell electrolyte and the sulfur compounds
  • PID dedicated for total VOC measurement.

 

Standards:

  • ISO 19739:2004, DIN 51855/7
  • ASTM D 7493-08
  • Correlation with EN13 725 / ASTM 679-E04

 

Features:

  • Automatic calibration/validation of the data, with embeded permeation tube for sulfurs. No span gases required.
  • Continuous monitoring with automatic online sampling
  • Analytical performances:
    • Specific, linear and very sensitive to sulfur compounds
    • Results validation by automatic standard injection
    • Long term stability using detector installed in reservoir
  • Extremely low maintenance
    • Very long life time detector with electrolyte, up to 10 years
    • Low gas consumption, can be reduced in option
    • More than 10 years data storage
    • No cylinders required thanks to internal calibration tube and gas generators
  • Automatic control with process device
  • Intelligents system with tunable and interactive alarm levels
  • PID detector in addition to MEDOR® detector
  • Powerful VISTACHROM Chromatotec® software:
    • Remote monitoring & injection control
    • Full traceability with on board archiving of results and chromatograms
    • QC Set up and control of threshold alarms
    • Data export by MODBUS / 4-20 mA / 0-10 V
    • Time stamp results